Features
l
Sensor has X-ray monitoring photodiode
l
Compactness
5.1 mm thickness excluding I/O connector part
l
High dynamic range: 12 bit
l
Long-term stability
For use under 100,000 shots
(60 kVp, 30 mR X-ray irradiation)
l
Resolution: 10 Lp/mm
l
1700 (H) × 1200 (V) pixel format
l
Pixel size: 20 × 20 μm
l
Coupled with GOS for X-ray imaging
l
100 % fill factor
l
Low dark signal
l
Low readout noise
l
MPP operation
l
AC/DC X-ray source adapted
Applications
l
General X-ray imaging
l
Non-destructive inspection
l
Intra-oral X-ray imaging in dental diagnosis
I M A G E S E N S O R
CCD area image sensor
S8986, S10128
Front-illuminated FFT-CCDs for X-ray imaging
S8986 is an FFT-CCD image sensor ideal for intra-oral X-ray imaging in dental diagnosis. S8986 has about 2 mega (1700 × 1200) pixels each of
which is 20 × 20 μm. S8986 delivers low dark current when operated in MPP (Multi Pinned-Phase) mode.
S10128 is an easy-to-use X-ray imaging module using S8986, with added functions such as a cable assembly and X-ray trigger circuit. S10128 is
pin compatible with S10127 (number of pixels is increased). S10128 is low cost type for S8985-02 and adapted both AC and DC X-ray sources.
1
I
Selection guide
Type No.
Cooling
Number of
total pixels
Number of
active pixels
Active area
[mm (H) × mm (V)]
S8986
S10128
Non-cooled
1708 × 1202
1700 × 1200
34 × 24
I
General ratings
Parameter
S8986
S10128
CCD structure
Fill factor
Number of active pixels
Pixel size
Active area
Vertical clock phase
Horizontal clock phase
Output circuit
Dimensional outline
Reliability
Window
Other
Full frame transfer
100 %
1700 (H) × 1200 (V)
20 (H) × 20 (V) μm
34 (H) × 24 (V) mm
2 phase
2 phase
Emitter follower without load resistance
39.5 (H) × 27.5 (V) mm
100,000 shots at 60 kVp, 30 m Roentgen
42.6 (H) × 30.6 (V) mm
GOS film
MPP mode (low dark current operation), module (S10128)