4
Specifications are subject to change without notice UDM40DS 040706
UDM40
RS422/RS485
(on request)
Module: BR SX
Bidirectional (static and
dynamic variables).
Display of data
reception/transmission
Multidrop, 2 or 4 wires,
1000 m
Directly on the module
by means of jumper
1 to 255, selectable
by means of key-pad
MODBUS RTU/JBUS
Serial output
LED
Connections
Distance
Terminalization
Addresses
Protocol
Data (bidirectional)
Dynamic (reading only)
Measurement, min value
max value
alarm status
All programming parameters,
min max reset
reset of latch alarm
8 data bit, no parity,
1 stop bit
selectable 4800, 9600,19200
and 38400 bit/s
By means of opto-couplers
4000 V
rms
output to
measuring inputs
4000 V
rms
output to
power supply input
(on request)
Module: BR SY
Bidirectional (static and
dynamic variables)
3 wires,
max. 15m
1 start bit, 8 data bit,
no parity, 1 stop bit
Selectable 4800, 9600,
19200 and 38400 bit/s
Same as RS422/485
(on request)
Over-range alarm,
up alarm,
down alarm,
down alarm with
start-up deactivation
up alarm with latch,
down alarm with latch
Adjustable from 0 to 100%
of displayed electric range
0 to 100% of displayed range
0 to 255 s
0 to 255 s
Selectable: normally energized
/de-energized
Three available coloures
(green, amber, red)
can be coupled to the alarm
status (traffic-light function).
500 ms, with filter excluded,
without alarm activation delay
1 with module BO R1
(relay output).
Static (reading/writing)
Data format
Baud rate
Insulation
RS232
Serial output
Connections
Distance
Data format
Baud rate
Other features
Alarm outputs
Alarm type
Alarm set-point
Hysteresis
On-time delay
Off-time delay
Output status
Display colour for alarms
Min response time
Output channels
2, independent with module
BO R2 (2 relay outputs).
4, independent with module
BO R4 (2 relay outputs +
2 open collector outputs).
BO R5 (4 relay outputs)
Type SPDT
AC 1: 8A, 250VAC
DC 12: 5A, 24VDC
AC 15: 2.5A, 250VAC
DC 13: 2.5A, 24VDC
Type SPST (NO)
AC 1: 5A, 250VAC
DC 12: 3A, 24VDC
AC 15: 1,5A, 250VAC
DC 13: 1,5A, 24VDC
4000 V
output to
measuring input,
4000 V
output to
power supply input.
NPN transistor type
V
ON
1.2 VDC/ max. 100 mA
V
30 VDC max.
By means of opto-couplers
4000 V
output to
measuring input
4000 V
RMS
output to
power supply input
(on request)
Module: BO AV
0 to 20 mADC, 0 to 10 VDC
Programmable within the
entire retransmission range;
allows to manage the
retransmission of all the
values from
0 to 20 mA / 0 to 10V
±0.2% FS(@ 25°C ±5°C)
≤
10 ms
± 200 ppm/°C
≤
700
≥
10 k
By means of opto-couplers
4000V
rms
output to
measuring input
4000V
rms
output to
power supply input
The two outputs cannot be
used at the same time.
(on request)
Relay output BO R1, R2, R4
Relay output BO R5
Insulation
Open collector output
Insulation
Analogue output
Range
Scaling factor
Accuracy
Response time
Termperature drift
Load: 20 mA output
10 V output
Insulation
Notes:
Excitation output
BQ LSE Module
Voltage
13 VDC ±10%
max. 50 mA
BQ LSF Module
Voltage
25 VDC ±10%
max. 25 mA
25V
rms
output to
measuring input
4000 V
rms
output to
power supply input
Insulation
Output specifications