KELLER
Specific Characteristics
Total Accuracy
(1)
Range 0…50°C
M
ANO
G
AUGE
/ R
ECORD
± 0,1% typ.
± 0,2% max.
M
ANO
C
ALI
0,05 % typ.
0,1% max.
Measuring Interval
(2)
sec
1
2
(1)
Includes linearity, repeatability, hysteresis, temperature and resolution of the display
(2)
x 2 for two transducers (not for M
ANO
C
ALI
)
Electrical Characteristics
Supply
Battery-Life
Output Signal (optional)
- Analog
(3)(M
ANO
R
ECORD
/ G
AUGE
)
- Numerical
(M
ANO
R
ECORD
/ C
ALI
)
Lithium-Battery 3,6 V (included)
> 200 days at continuous operation
(4)
4..20 mA or 0..10 V
RS 232
(3)
The use of an analog signal requires an external supply between 10…28 VDC
(4)
> 100 days for M
ANO
C
ALI
Options
(on request)
- Manometer Housing in Metal*
- 1 or 2 external Transducers (remote)*
- Transmitter Output Connector*
- Special Programs
- Logger Software and K101 Cable
- Intrinsically Safe Version (EEx-ia)*
*
only M
ANO
R
ECORD
and M
ANO
G
AUGE
LOGGER
The Logger-Software allows the read-out of the stored
data from the Intelligent Manometer. The data transfer
takes place via the serial interface on the PC.
The M
ANO
R
ECORD
/ M
ANO
C
ALI
is connected through a
cable K101 (supplied with Logger software).
The LOGGER-software offers many further processing
possibilities of the read-out data, and meets all the
requirements of modern Windows-Software:
- Graphical Display
- Tabulated Display
- Export into other Windows-Programs
- Storage into Files
Mechanical Characteristics
Manometer Housing
Transducer
Total Weight
Pressure Connection
Protection
Polyamid 12
See data sheet of transducer
approx. 250 g
See data sheet of transducer
IP 65
Environmental Characteristics
Operating Temperature
Compensated Range
(5)
Storage Temperature
Display
(5)
°
C
°
C
°
C
°
C
- 10
…
60
0
…
50
- 30
…
80
0
…
50
(5)
others upon request
Vented Gauge or Absolute
Range
30 mbar
300 mbar
1000 mbar
-1
…
2 bar
10 bar
30 bar
Resolution
0,01 mbar
0,1 mbar
0,3 mbar
1 mbar
3 mbar
10 mbar
Sealed Gauge or Absolute
Range
Resolution
100 bar
300 bar
600 bar
1000 bar
0,03 bar
0,1 bar
0,2 bar
0,3 bar
In M
ANO
G
AUGE
and M
ANO
R
ECORD
the
measuring ranges are divided by 12 bit
or 3000 points. In M
ANO
C
ALI
over 13
bit or 6000 points.
The resolution is determined by the
measuring range and the number of
measuring points.
SPECIFICATIONS
M
ANO
G
AUGE
/ M
ANO
R
ECORD
M
ANO
C
ALI
Sealed or Vented Gauge
Range
-1
…
2 bar
-1
…
10 bar
-1
…
25 bar
200 bar
350 bar
700 bar
Resolution
0,5 mbar
1,8 mbar
4,3 mbar
0,03 bar
0,05 bar
0,1 bar