Libertyville, IL (847) 680-7400
MV
60
Specifications
Maximum leakage current, each
line-to-ground
@ 120VAC 60 Hz:
.07mA
@ 250VAC 50 Hz:
.13mA
Hipot rating (one minute):
line-to-ground
2250 VDC
line-to-line
1450 VDC
Operating frequency:
50/60 Hz
Rated voltage:
120/250 VAC
Rated current:
@ 120VAC
@ 250VAC
3MV
3A
2.5A
6MV
6A
5A
10MV
10A
8A
20MV
20A
16A
Minimum insertion loss in dB:
Line-to-ground in 50 ohm circuit
Current
Frequency-MHz
Rating
.15
.5
1
2
5
10
20
30
3A
14
19
20
30
46
40
34
31
6A
19
27
30
38
50
40
35
10A
15
25
26
34
46
50
44
42
20A
18
30
34
46
40
36
20
Line-to-line in 50 ohm circuit
Current
Frequency-MHz
Rating
.15
.5
1
2
5
10
20
30
3A
33
65
60
53
50
6A
40
65
60
57
55
10A
33
65
55
20A
25
65
60
57
45
MV Series
MV Series
The MV series answers the need for a medical filter to help
equipment comply with increasing noise sources in
medical applications while still meeting leakage current
requirements of UL2601.
Less strict than previous UL medical standards, UL2601
limits leakage current for patient care equipment. Due to
the increased need for higher performance filtering, UL has
allowed more leakage current than in previous standards.
The MV takes advantage of this to offer improved line-to-
ground performance in a moderately priced filter. Higher
line-to-line performance is achieved through its additional
differential mode rod inductors. Overall, the filter provides
a good solution to emission and immunity problems, at a
moderate cost, in a reasonable package size.
Electrical Schematic
Multi-purpose Medical Filter
for Power Line Noise Protection
with Limited Leakage Current
UL Recognized
CSA Certified
VDE Approved
Resistor location for reference only.
3A, 6A, 10A